produkt zdjęcie
TECHNICAL

 

Water treatment systems with an industrial capacity of 100 to 1000 dm3 / h. The stations allow receiving purified water of the third* purity class according to ISO 3696: 1999. Each device is designed and manufactured to meet individual customer needs. These devices can be used in the industry (pharmaceutical, food, cosmetics, electronics, etc.) as a central unit producing demineralized water. The demineralizer, by distributing an appropriate network, can supply several laboratory rooms or floors in the building, as well as point laboratory devices.

* depends on feed water quality, retention rate is 96-99%

ikona

Technical 100

Cat.no. DT-0100-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 100 – 150 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian, Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾” or 1″,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

Technical 200

Cat.no. DT-0200-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 150 – 250 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾” or 1″,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

Technical 300

Cat.no. DT-0300-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 250 – 350 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

Technical 500

Cat.no. DT-0500-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 400 – 600 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾” or 1″,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

Technical 700

Cat.no. DT-0700-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 700 – 800 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

Technical 1000

Cat.no. DT-1000-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 900-1100 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

• UV lamp (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

– Purified water in the device meets the requirements of ISO 3696: 1999 for waters of the third * degree of purity.

* depends on the quality of the feed water, the degree of retention is 96-99%

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
produkt zdjęcie
SPRING

 

Water treatment systems with an industrial capacity of 100 to 1000 dm3/h. The stations allow obtaining high purity water that meets the requirements of ISO 3696: 1999, ASTM for first, second and third degree purity waters and in accordance with FP.  Each device is designed and manufactured to meet individual customer needs. These devices can be used in the industry (pharmaceutical, food, cosmetics, electronics, etc.) as a central unit producing demineralized water. The demineralizer, by distributing an appropriate network, can supply several laboratory rooms or floors in the building, as well as point laboratory devices.

* depends on feed water quality, retention rate is 96-99%

ikona

SPRING 100

Cat. no. DS-0100-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 100 – 150 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (column 25 dm3 capacity),

• UV lamp 185 / 254nm or 254 nm (option),

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

SPRING 200

Cat. no. DS-0200-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 150 – 250 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (column 25 dm3 capacity),

• UV lamp 185 / 254nm or 254 nm (option),

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display (option),

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

SPRING 300

Cat. no. DS-0300-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 250 – 350 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (column 25 dm3 capacity),

• UV lamp 185 / 254nm or 254 nm (option).

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

SPRING 500

Cat. no. DS-0500-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 400 – 600 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (2 columns 25 dm3 capacity each),

• UV lamp 185 / 254nm or 254 nm (option),

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

SPRING 700

Cat. no. DS-0700-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 700 – 800 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (2 columns 25 dm3 capacity each),

• UV lamp 185 / 254nm or 254 nm (option),

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

Files to download
ikona

SPRING 1000

Cat. no. DS-1000-0C

 

Technical specifications:

– The device is supplied with tap water.

– Degrees of water purification:

• filtration on sediment filters:

cascade filtration on 20µm and 5µm high efficiency sediment filters,

• filtration on carbon filters:

filtration on granular activated carbon to remove organic compounds, chlorine and chlorine derivatives,

• softening process (option):

automatic softening station – rinsing and regeneration of the bed is carried out automatically, compact construction – corrosion resistant ion exchange column (fiberglass tank) placed inside the salt casing, high capacity for removing hardness ions,

• reverse osmosis station:

efficiency: 900 – 1100 dm3 / h (depending on the model), retention ratio 96-99%, recovery rate 60%, high pressure pump, retentate and permeate rotameters,

            • demineralization on a mixed ion-exchange bed (2 columns 25 dm3 capacity each),

• UV lamp 185 / 254nm or 254 nm (option),

• 0.45 / 0.2µm microfiltration capsule (option).

– Automatic and maintenance-free operation of the device.

– Retention rate is 96-99%.

Conductivity of purified water <0.06 µS / cm.

– Water intake point – third * purity class according to ISO 3696: 1999.

– Possibility of installing an additional water intake point – first class of purity according to ISO 3696: 1999 and in accordance with FP.

– Possibility to install an additional water intake point – third purity class according to ISO 3696: 1999.

– Tank for storing purified water (capacity to choose).

– Automatic system shutdown when the tank is full or the water intake is closed.

– Can be connected to a dishwasher, autoclave, analyzer, etc.

– Possibility of creating a water distribution network with intake points covering several rooms or floors in a building.

– Control water intake points.

– Automatic membrane rinsing (possibility of individual setting of the period and time of membrane rinsing).

– Forced flushing of membranes (service).

– System designed for cold water supply: 5-40 ° C.

– Possibility of self service by the User (without having to call the service).

– Power supply: 230V / 50Hz.

– Stainless steel frame.

 

Functions monitoring system operation:

– The device is equipped with a microprocessor control and measuring system having:

• color graphic display with Touch Panel function,

• conductivity meter measuring the conductivity and temperature of tap water,

• conductivity meter measuring the conductivity and temperature of purified water after reverse osmosis,

• conductivity measurement in µS / cm or MOhm units,

• automatic temperature compensation,

• continuous control and preview of the degree of retention (degree of retention) of RO membranes,

• clock displaying date and time,

• alarm informing about mechanical and carbon filter replacement,

• alarm informing about replacement of the RO module,

• alarm informing about exchange of ion exchange bed,

• alarm informing about the UV lamp radiator replacement (option),

• information about the tank filling level on the device display,

• alarm informing about the replacement of the microfiltration capsule (option),

• preview of service dates,

• menu in Polish, English, Russian or Spain on display,

• built-in RS 232 interface for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• built-in USB connector for communication with a computer ensuring the possibility of individual adjustment of service frequency and alarm levels,

• computer program enabling individual settings of alarm thresholds and data archiving,

          • built-in manometer.

 

Functions securing the system operation:

– Interruption of system operation with:

• low feed water pressure (no feed water),

• a full tank / closed water intake point.

– Thermal protection of the osmotic module, automatic stopping of the system operation at the supply water temperature below 40C or above 400C.

– Ability to stop the system when any alarm occurs.

– System autostart capability.

– Preview of monitoring messages / alarms.

 

Purified Water Parameters:

Purified water in the device meets the requirements of ISO 3696: 1999 for first, second and third degree water.

The water obtained meets the microbiological and physicochemical requirements of FP for purified production water *.

* system equipped with a UV lamp and a microfiltration capsule

 

Additional equipment / services:

– tank for storing purified water:

• pressure – option: 80 l, 110 l, 230 l, 320 l or 450 l,

• pressureless – capacity to be agreed (on request),

– compact device housing made of stainless acid-resistant steel (all system components, except for the tank, installed inside the housing),

– automatic softening station,

– tank housing made of stainless acid-resistant steel,

– adjustment of conductivity value from 0.06 µS / cm to 15 µS / cm,

– purified water recirculation,

– full DQ, IQ, OQ, PQ qualification procedure with documentation,

– cooperation with building management systems (BMS).

 

Required connections at the installation site:

– cold tap water connection ¾ “or 1”,

– drain to sewage system (sewage grate),

– 230V socket.

 

Files to download